EXPERIMENTAL PAPRAMETERS
Linear range
The calibration curve of Figure 1 shows a linear range to at least 12 mM of lactate.
LIMIT OF DETECTION
The Limit of Detection (LOD) for this measurement using Photopette® is determined by factoring in the standarddeviation for blank measurements as well as experimental data using the equation given below:
LOD= 3 x SD blank /Slope standard curve.
Standard Deviation for blank measurements with 50 repeats using same CuveTip® was found to be 0.001 AU. Thus, the limit of detection for the lactate assay with Photopette® using the equation above was 0.024 mM
LOD = 3 x 0.001 AU / (0.125 AU/mM)
LOD = 0.024 mM
LACTATE CONCENTRATION OF UNKNOWN SAMPLES
An undiluted serum sample was measured and the average of triplets resulted in an absorbance of 0.102. The linear regression of the lactate standards of Figure 1 is:
y = 0.125x + 0.056
Rearrangement to the lactate concentration “x” results in:
x = (y – 0.056) / 0.125 Inserting the absorbance of the unknown sample results in:
x = (0.112 – 0.056) / 0.125
x = 0.448 mM lactate
EXPERIMENTAL VARIATIONS
The same measurement procedure can be used to determine lactate concentrations in other samples such as various beverages, food and biological samples including serum, plasma and fermentation media. For strongly colored samples the absorbance at 340 nm must be taken into account and corrected for. Non-liquid and turbid samples need extraction and/or filtration before the assay can be performed.
LIMITATIONS
The calibration-range was limited by the concentration range recommended by Sigma-Aldrich. Therefore, the range of standard curve is not a limitation of the Photopette® device.